Now, (Hmin.) (the minimum variance per unit length of the eluted peak) to the first approximation, has been shown (see book 9 of this series) to be,

                                      (40)

where (dp) is the particle diameter of the packing.

The length (L) of the column is ,

                                 L = nHmin.                                                             (41)

Substituting for (n) from (39) and for (Hmin.) from (40) in (41),

                        (42)

  

Similarly, it has been shown (15), the optimum mobile phase velocity is given by,